Olympus U-FWT Instrukcja obsługi - Strona 9
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Mikroskop Olympus U-FWT. Olympus U-FWT 12 stron. Transmitted light/reflected light/observation filter wheels
@
²
Fig. 3
ƒ
...
|
Fig. 4
2
Mounting the Filter Wheel on the Microscope
U-FWT Filter Wheel (Figs. 3 & 4)
|
}The U-FWT can be mounted only to a BX2 series microscope (BX41/51/
52/61/62 microscope frame compatible with reflected light illmination).
1. Loosen the clamping screw ² of the lift prevention pin @ using the Allen
³
screwdriver.
2. Loosen the 4 clamping screws ³ on each side using the Allen screw-
driver.
3. Fit the filter wheel mount | on the filter holder on the BX2 series micro-
scope.
4. While pushing down the filter wheel mount, tighten the clamping screws
³ uniformly so that the filter wheel mount is parallel with the microscope
base.
5. While pushing down the lift prevention pin @, tighten the clamping screw
².
6. Fit the filter wheel ƒ on the filter wheel mount |. During this, place the
connection cable ... in the cable groove on the back side of the filter
wheel ƒ.
7. Tighten the clamping screws on the left side panel of the filter wheel using
the Allen screwdriver.
8. Connect the connection cable ... of filter wheel to one of connectors FW1,
2 and 3 on the BX-UCB.
U-FWT/FWR/FWO
(Figs. 3 - 5)
6