Olympus HysteroFlow Skrócona instrukcja obsługi - Strona 6
Przeglądaj online lub pobierz pdf Skrócona instrukcja obsługi dla Sprzęt medyczny Olympus HysteroFlow. Olympus HysteroFlow 8 stron. Hysteroscopic fluid management system
Warning Message
Change Fluid Bag
Deficit Reached
Overpressure
Deficit Exceeded
Critical Overpressure
Rapid Deficit
6
HYSTEROFLOW QUICK REFERENCE GUIDE
All manuals and user guides at all-guides.com
B E G I N T H E P R O C E D U R E
W R N I N G S
Alert
Cause
3 Beeps
> 500 ml remaining
3 Beeps
Deficit exceeds the preset value
3 Beeps
Actual pressure 10 mmHg above nominal
pressure
10 Beeps
In increments of 100 ml above the preset
value
10 Beeps
Actual Pressure > 200 mmHg above nominal
pressure
10 Beeps
Deficit change > 300 ml in 1 minute