Chapter 6 Troubleshooting
6.2

Returning the colposcope for repair

28
Irregularity
description
Focus cannot be
adjusted.
The focus range is too
narrow or wide.
The entire visual field
is hard to observe
(when the operator is
wearing glasses).
The arm moves on its
own.
The microscope body
tilts under its own
weight, or tilting is
heavy.
Olympus is not liable for any injury or damage which occurs
as a result of repairs attempted by non-Olympus personnel.
Before returning the colposcope for repair, contact Olympus. With the
colposcope, include a description of the malfunction or damage and the name
and telephone number of the individual at your location who is most familiar with
the problem. Also include a repair purchase order.
COLPOSCOPE OCS-500
Possible cause
The limit of the focus adjustment
range has been reached.
The limit of the zoom range has
been reached.
The height of the eyepiece
shade(s) is not properly adjusted.
The vertical balance has not
been adjusted.
The rotation tension level is too
low for each rotating part.
The tilt tension has not been
adjusted.
Solution
Move the microscope
body up or down so that
observation is possible.
Use a magnification within
the current zoom range.
Adjust the height of the
eyepiece shade (Section
4.7).
Adjust the balance.
Adjust the rotation
tension.
Adjust the tilt tension.