Inspection of the CVP counter
When this instrument is connected to the following
equipment, perform inspection of the CVP counter.
– EVIS EXERA II video system center (CV-180, CV-
160)
– VISERA Pro video system center (OTV-S7Pro)
1
Display the output image from the video system
center on the monitor by referring to the instruction
manuals for the video system center and the
monitor.
2
Confirm that the monitor screen displays the "CVP
counter" (see Figure 3.20).
CVP counter
3
3
If the "CVP counter" is not displayed on the
monitor screen, refer to "Irregularity: The CVP
counter is not displayed" in "7.1 Troubleshooting
Guide" in "INSTRUCTIONS (ADVANCED)."
24

3.7 Check the Printed Picture

Figure 3.20
3.7 Check the Printed
Picture
Check whether the printed picture is desired.
1
Observe, for example, the palm of your hand using
the endoscope, while referring to the instruction
manual of any connected equipment.
2
Capture a picture, while referring to the instruction
manual of any connected equipment.
Confirm that the number displayed on the "CVP
counter" is increased by one. If the number does not
increase, refer to "Irregularity: The number
displayed on the CVP counter does not increase
even if an image has been captured" in "7.1
Troubleshooting Guide" in "INSTRUCTIONS
(ADVANCED)."
3
Print.
4
Check that the printed picture is of a suitable
quality.
If it is not of the desired picture quality, adjust the
picture quality, referring to "Chapter 5 System
Setup" in "INSTRUCTIONS (ADVANCED)."
After adjustment, print and check again.