Fig. 24
REFLECTED FLUORESCENCE SYSTEM
Fuse replacement (Fig. 24)
# Set the main switch to "
cord before replacing fuses.
1. Remove each of the fuse holders 9 by turning it counter-
clockwise using a flat-blade screwdriver and pulling out.
2. Replace both fuses with new ones.
# Always use the designated fuses. Otherwise, a fire hazard
may result.
Applicable fuses
" (OFF) and unplug the power
U-RFTL50-100:
T4A(H)250V
(LITTELFUSE 215004), 2 pieces
U-RFTL50-200:
T2.5A(H)250V
(LITTELFUSE 215025), 2 pieces
FOR CKX41
28