Olympus CKX41 Instrukcja naprawy - Strona 5
Przeglądaj online lub pobierz pdf Instrukcja naprawy dla Sprzęt laboratoryjny Olympus CKX41. Olympus CKX41 43 stron. Reflected fluorescence system
Również dla Olympus CKX41: Przegląd (7 strony), Broszura i specyfikacje (4 strony), Instrukcja obsługi (36 strony)
CKX31/CKX41
(7) Attaching a sub-stage(CK2-SS) to the plane stage, the necessary stage area can be set
and working space can be secured easily.
CKX31
(1) Observation is possible in F.N.20.
CKX41
(1) A low-priced tilting obserbation tube is provided to facilitate observation in the standing
or sitting position.
(2) Fluorescence observation can be made by mounting a reflected light fluorescence attachment
(B-excitation, G-excitation). As option, it is applicable to U-excitation.
(3) The digital camera(DP12) can be attached to the trinocular tube, as can various types of
video camera.
(4) Observation is possible in F.N.20 to 22.
3. Using conditions
CKX31/CKX41
(1) Ambient temperature:
Temperature: 0 - 40 degrees C(32-104 degrees F), Humidity: 30-90%
(2) Range of combination: As indicated by the CKX41/CKX31 system chart.
(3) Applicable objective
1) Phase contrast observation:
Applicable slider
Applicable objective
*If a combination of slider and objective is changed, phase contrast effectiveness
can not be obtained.
*In a combination of IX2-SL, objective other than the above appicable objective is possible
to use as long as the objective corresponds to PHL/PHC/PH1/PH2.
2) Brightfield observation:
PLC4X, PLC10X, ACH20X, LCPLFL40X
*Since illumination does not cover the N.A. of LCPLFL40X objective, resolution is not
fully obtained in this objective.
*In brightfield observation, objective other than the above ones can not be used.
If used, visibility is deteriorated because N.A. of objective is larger than the above
applicable objective.
(4) Applicable container sizes
When a ULWCD is mounted (phase contrast observation):
70mm or less in overall height
When no ULWCD is mounted (brightfield observation):
148mm or less in overall height
CKX31
(1) Since the observation tube is fixed with the frame, it can not be changed.
CKX41
(1) Applicable observation tube
F.N.20: U-CBI30-2, U-CTR30-2, CKX-TBI
F.N.22: U-TBI-3, U-TR30-2, U-BI30-2
(2) Applicable eyepiece
WHB10X/10X-H(F.N.20), WH10X/10X-H(F.N.22)
(when WHC10X is used, visibility on the perphery is not clear.)
A. OUTLINE OF PRODUCT
IX2-SLP
UPLFL4XPH (PHL)
UPLFL4XPHP
CACH10XPHP
CPL10XPH(PHC), PL10XPH(PH1)
LCACH20XPH(PHC), LCPLFL20XPH(PH1)
LCACH20XPHP
LCACH40XPHP
LCACH40XPH(PH2), LCPLFL40XPH(PH2)
A-2
IX2-SL
2