Attaching the Eyepieces
Fit and insert gently an eyepiece into each eyepiece sleeve.
# Both the left and right eyepiece sleeves have the eyepiece posi-
tioning grooves ². Be sure to fit the positioning pin @ of the eye-
piece into the positioning groove ² of the eyepiece sleeve when
mounting each eyepiece.
7
Attaching the Halogen Bulb
}The applicable halogen bulbs are the 12V100WHAL-L long-life bulb
(PHILIPS 7724) and the 12V100WHAL high-intensity bulb (PHILIPS 7023).
1. Using the Allen screwdriver provided with the microscope, fully loosen
the clamping screw @ on the top of the lamp housing.
2. Lift the lamp housing ² to remove.
3. Tilt the lamp socket ³ by 90° in the direction of the arrow.
4. While holding down the lamp clamping lever |, hold the halogen bulb
ƒ by means of a piece of gauze, and insert the pins ... all the way into
the pin holes †.
Return the lamp-clamping lever to the original position to clamp the bulb.
³
Do not touch the bulb directly with bare hand. If fingerprints are
attached on it, wipe thoroughly with a soft cloth to prevent the ser-
vice life from dropping and the bulb from cracking.
5. Fit the lamp housing ² from above and tighten the clamping screw @
while pushing down the screw by force. (Fig.12)
Bulb replacement during use or right after use
The bulb. lamp housing and areas near these will be extremely hot
...
during and right after use.
After setting the main switch to "
power cord from the power outlet, allow the old bulb and lamp
socket to cool before replacing the bulb with a new one of the
designated type.
MX61A
(Fig. 11)
(Fig. 12 to 14)
" (OFF) and unplugging the
16