TROUBLESHOOTING GUIDE
Under certain conditions, performance of this equipment may be adversely affected by factors other than defects. If a problem
occurs, please review the following list and take remedial action as needed.
If you cannot solve the problem after checking the entire list, please contact your local Olympus representative for assistance.
Problem
Field of view is not evenly illuminated.
Color reproduction in the photography is
poor.
Contrast is poor.
The view is noticeably different between
the left and right fields of view.
11
Cause
The light guide is not mounted properly.
The bulb of the light source unit is not
mounted properly.
The brightness control of the power trans-
former is not set to the maximum posi-
tion.
The filters (30.5S-LB80 and 30.5-LBD) are
not used (when a daylight type film is in
use).
The quarter-wave plate is not adjusted.
The quarter-wave plate clamping posi-
tion is not adjusted properly.
Remedy
Mount the light guide properly.
Mount the bulb properly.
Set the brightness control to the maxi-
mum position.
Mount the filters (30.5S-LB80 and 30.5-
LBD) on the LG-PS2 light source unit.
Adjust the plate. (Page 9)
Adjust the clamping position correctly.
(Page 9)