Olympus U-UCD8A Instrukcja obsługi - Strona 11

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Sprzęt laboratoryjny Olympus U-UCD8A. Olympus U-UCD8A 40 stron. Motorized universal condenser

Olympus U-UCD8A Instrukcja obsługi
Aperture iris
diaphragm
image
Objective pupil
Fig. 1
Field Iris Diaphragm
70-80%
{The field iris diaphragm controls the size of the illuminated
30-20%
area. By stopping down the field iris diaphragm, in accordance
with the objective in use, until its image circumscribes the
field of view, stray light can be reduced, which in turn increases
the definition and contrast of the image.
Aperture Iris Diaphragm
{The aperture iris diaphragm controls the numerical aperture
(N.A.) of the illuminator. In order to achieve the optimum objec-
tive performance, the opening of the aperture iris diaphragm
should be matched with the N.A. of the objective in use. This
will result in better image contrast and resolution as well as
increased depth of focus.
{As microscopic specimens are usually low in contrast, reduc-
ing the diaphragm opening to 70% or 80% of the objective's
N.A. will generally provide an image of acceptable quality. To
check the opening, after completing focus adjustment, remove
one of the eyepiece lenses and look into the empty eyepiece
sleeve. As you stop down the aperture iris diaphragm, the iris
diaphragm image can be seen in the objective pupil. (Fig. 1)
U-UCD8A
8