Problem
g. No interference color appears
during Nomarski observation.
h. Interference color appears
during Nomarski observation
but color is uneven.
21
Cause
Polarizer is not inserted in light path.
Analyzer is not inserted in light path.
DIC prism is not inserted in light path.
DIC slider is not inserted in light path.
Polarizer and analyzer are not in "crossed
Nicol" position.
Vertical positioning of condenser is in-
correct.
Incorrect optical element is inserted in
light path.
Incorrect objective is in use.
Remedy
Engage polarizer in light path.
Engage analyzer in light path.
Engage DIC prism by rotating turret.
Engage DIC slider in light path.
Re-adjust polarizer.
Center condenser.
Engage optical element that matches
objective in use by rotating turret.
See tables in "Optical Elements and
Compatible Objectives" on pages 26 to
29 and use correct objective.
Page
13
13
14
13
13/14
35
26-29
26-29