AM1710 Secondary Reference PRT User's Guide

2 Specifications

2.1 Specifications
Temperature Range
Resistance at 0 °C
Temperature Coefficient
Accuracy
Drift
Short Term Stability
Thermal Shock
Hysteresis
Self-heating
Response Time
Measurement Current
Sensor Length
Sensor Location
Insulation Resistance
Sheath Material
Dimension
External Leads
Handle Dimension
Handle Temperature Range
Calibration Options
Rev. 201803V1
-60°C to 160°C
Nominal 100 Ω
0.003925 Ω/ Ω/°C
±0.015°C at 0.01 °C
±0.01 °C at TPW after 100 hours at 160 °C
±0.007°C
±0.005°C after 10 times thermal cycles from minimum to
maximum temperatures
<=0.005°C
50 mW/°C
9 seconds for 63% response to step change in water moving
at 3 feet per second
0.5 mA or 1 mA
32 mm
5 mm from tip
>1000 MΩ at room temperature
tm
Inconel
1710-12: 0.25 inch X 12 inch (6.35 mm X 305 mm)
1710-9: 0.187 inch X 9 inch (4.75 mm X 229 mm)
tm
Teflon
–insulated copper wire, 4 leads, 2.5 meters
1710-12: 15mm (OD) x 65mm (L)
1710-9: 10mm (OD) X 50 mm (L)
-50°C to 160°C
NIST traceable calibration and data, Ordering # 5017;
ISO17025/NVLAP Accredited Calibrations available per
request