Operating Modes and Displayed Information
Single Polarity (Positive or Negative) Operating Mode
Selected Polarity
Peak Ion Count
System Warning
See page 9 for details.
Scales:
X10³ (multiply ion count by 1000), Range = 10 ions/cc to 2 million ion/cc
X10⁶ (multiply ion count by 1 million), Range = 1000 ions/cc to 200 million ions/cc
This mode allows for the alternating measurement of each polarity to be displayed. The SUM
feature adds the total number of positive and negative air ions and displays the cumulative count.
Ion polarities are alternately updated every 7.5 seconds.
The rotating timer indicates the update cycle.
Temperature
°C/°F
Ion Count Averaging
Dual Polarity Operating Mode
Rotating Timer
Humidity
Real-Time Ion Count
Scale X10³ or X 10⁶
MicroSD Card Installed
Positive Ion Count
Negative Ion Count
SUM Ion Count
5