5. REPROCESSING

The following step-by-step instructions describe the reprocessing procedure for the
DrillCover. It is important to carefully follow these steps to ensure the device is not damaged
and remains safe to use.
Separate DrillCover Linen from hardware and other sharp and hard objects before
reprocessing.
Reprocessing linen together with sharps and hard objects may result in damage to the
linen and compromised sterility.
The lifetime of the DrillCover linen is 75 uses when following these guidelines.
5.1. Wipe Drill
1. Clean and disinfect drill and battery with germicidal cloth. Follow the manufacturers
instruction for contact time.
5.2. Clean DrillCover
Ensure the linen is reprocessed separately
from the Chuck Adapter.
If the disassembly
step 8 in Section 4.8.
not followed by the sterile team member,
disassemble the chuck adapter from the
drill cover by turning in counter clockwise
direction.
1. Immerse chuck key, chuck adapter and
linen into a freshly prepared solution of
neutral enzymatic solution designed for
surgical instruments (ex. Medzyme) and
allow to soak for 5 minutes.
was
DrillCover Hex System: Instructions For Use
13