fiagon E 01 1905 Instrukcje użytkowania - Strona 3
Przeglądaj online lub pobierz pdf Instrukcje użytkowania dla Adapter fiagon E 01 1905. fiagon E 01 1905 4 stron. Flexarm adapter for navigation sensor for side rails 25 x 10 mm
Instructions for Use
3
Mounting of Navigation Sensor Flexarm
To attach the Navigation Sensor Flexarm to the rail of 25 x 10 mm please follow the installation steps:
I.
Attach the rail clamp of the device to an appropriate position on the rail.
II.
Fix the clamp firmly with the screw.
III.
Adjust the navigation sensor to the surgical area
and fix its position by tightening the articulating
arm`s knob.
IV.
Insert the plug of the sensor into the appropriate socket "sensor" (metal socket) on the navigation
unit. Note that the red marking on the plug is in the 12 o'clock position.
CAUTION: Using too much force while trying to plug in the sensor cable in a different position
may cause damage to the plug and/or the socket.
After usage / disassembling
For disassembling please perform installation steps in reverse order.
Cleaning instructions
The Navigation Sensor Flexarm is a reusable component.
CAUTION: The navigation sensor or parts of the sensor must not be sterilized.
Clean the Navigation Sensor Flexarm by hand and disinfect using an aldehyde-free disinfectant.
CAUTION: Do not immerse the sensor, or parts of the sensor in the disinfectant or cleaning agent
Preventive Maintenance Instructions
Maintenance is done by authorized personnel only. Recommended maintenance period is 1 year and includes
accuracy testing according to a described procedure.
Copyright © 2018
Products
Fiagon GmbH
E 01 1905
DocumentID
Version
DOC-18-00044
2.00.0000
Date
3 / 4