Charging, cleaning, and disposal
Charging the AV500
Verify the charging cradle is properly connected (see the
Setup section). Place the AV500 in the charging cradle with
the LCD facing outward to begin charging. A battery charge
icon will display when the AV500 is charging. The light on the
charging cradle will only turn on when a device is detected.
If the light turns on when there is no device placed in the
charging cradle please call AccuVein support for assistance.
8
Operate the device in
powered hands-free
stand
80% to 100% charge
60% to 80% charge
40% to 60% charge
20% to 40% charge
Inoperable due to
low charge
To ensure the battery remains charged and ready for use,
you can store the AV500 in its charging cradle or a powered
AccuVein hands-free accessory. The AV500 battery cannot
be overcharged when approved charging devices are used.
CAUTION: Use only the AccuVein P510 power supply with
the CC500 cradle or other AccuVein approved accessories.
When the AV500 battery gets low the battery indicator will
flash yellow or red and beep repeatedly. Before the battery is
completely depleted, an alert requesting to charge the bat-
tery will be displayed on the screen. The AV500 will resume
normal operation once it has been charged.
In normal operation, the AV500 battery is expected to be
replaced about every two or three years. Battery life depends
on factors such as usage times, temperature of the device,
and number of charging cycles. Longer charge cycles or
shorter operating periods are indications that the battery
should be replaced.
To replace the battery the device can be returned to an AccuVein
approved facility approved to perform a battery replacement.
The AV500 requires no periodic maintenance, field calibration,
and is not field repairable.