Assembly/installation
4.2 Laser gas analyzers SITRANS SL and LDS 6
4.1
Gas analyzers OXYMAT 6/OXYMAT 61/OXYMAT 7
NOTICE
Magnetism
Devices of type OXYMAT 6 or OXYMAT 61 or OXYMAT 7 emit magnetic stray fields. This is why
reason magnetically sensitive devices should not be installed in the immediate vicinity of these
devices.
• Keep a distance in such cases. This distance can be up to 50 cm depending on the sensitivity
4.2
Laser gas analyzers SITRANS SL and LDS 6
Damage to the optics
If sensors are to be purged, they must not be mounted until the purging medium is available
and connected.
• Never switch off the purge function after purging, otherwise the wedge optics can easily be
12
of the affected device.
CAUTION
damaged.