3.2
Electrical safety
SIE
Cables and plugs
•
Check visible system cables and plugs for damage.
SIE
Fluoroscopic timer
•
Check: See section on "Compulsory radiation switch off".
SIE
Audible warning signal
•
Check: See section on "Compulsory radiation switch off".
SIE
Radiation indicators
•
Check: See section on "Compulsory radiation switch off".
SIE
Radiation release switch
•
Check: See section on "Compulsory radiation switch off".
SIE
Compulsory radiation switch off
Test the functioning of the audible warning signal and compulsory radiation switch off (if
required) according to the country-specific regulations. Also perform the following
checks:
•
X
Release fluoroscopy for one minute. Use a wristwatch to check the proper functioning
of the fluoroscopic timer.
•
All radiation indicators on the system operating console and monitor trolley must illumi-
nate during fluoroscopy release.
•
X
Check the function of the radiation release switches (hand and foot switches).
•
Check the radiation release switches for damage.
•
Check the cables of the hand and foot switches for mechanical damage.
•
Move the cables of the hand and foot switches to check for breaks and short circuits.
NOTE
SIE
Dose rate
•
Switch the system on
•
X
Check the dose rate (refer to the service instructions for SIREMOBIL ISO-C).
SIE
Iris collimator
SIREMOBIL Iso-C
•
Depending on the programming, the audible warning signal
has to sound 4.5, 5 or 9.5 minutes after the fluoroscopy has
started. It sounds again every 5 or 10 minutes. It is turned off
by pressing the -0- button once.
•
Pressing this button again resets the fluoroscopic timer to "0".
•
Depending on the country-specific programming, either the re-
quired radiation shut off does not occur or it occurs every 5 or
10 minutes and then every 5 or 10 minutes after that.
SPR2-230.101.02.06.02
07.05
Safety Inspection
Page 32 of 46