Operating (hardware)
3.4 Verificator software - sensor - magnetism
Is only displayed if during verification some data relating to the magnetic properties deviates
from the data stored in the SENSORPROM
See On site troubleshooting (Page 33)
Press ESC to cancel the test.
E3:
Magnetism fault
Press ESC to cancel
If the test fails you can get the following messages
Is displayed if the magnetic circuit is open.
Verification finish
Sensor:
Open circuit
Press GO to continue
Is displayed if the measured values are out of range.
Verification finish
Sensor:
Out of range
Press GO to continue
Is displayed if the gathered readings are unstable.
Verification finish
Sensor: Selftest Fail
Unstable reading
Press GO to continue
Magnetism test finished successfully, sensor passed the test.
I3:
Magnetism OK
Please wait .....
14
unit.
®
Sensor Failed
Selftest Fail
Selftest Fail
Sensor passed
SITRANS F M Verificator
Operating Instructions, 07/2019, A5E45134218-AB