Shields are common, but not grounded to chassis. Run cable shields through SHLD terminals
♦
and ground at Integrator only.
Note
Ensure that the connection between TB-2 and TB-17 is made.
SITRANS WFS300
Operating Instructions, 03/2021, A5E32880243-AC
5.2 Unit with sensing head mounted LVDT conditioner card
Connecting
17