Crowcon C-Test Podręcznik użytkownika - Strona 4
Przeglądaj online lub pobierz pdf Podręcznik użytkownika dla Sprzęt testowy Crowcon C-Test. Crowcon C-Test 12 stron. Gas (bump) test and calibration station for use with the crowcon clip and crowcon clip+
Crowcon C-Test User Manual
2.1.
Indications
No USB/USB Fail Indication: If the USB is properly installed the power light will remain solid during
testing. A blinking power light during testing indicates either no USB or a USB failure. A blinking light
also indicates the tests are not being recorded.
Pass/Fail Indication: A successful test will result in a
Once the Power button has displayed the final result
detectors.
It is important to ensure that the 'TEST' light is not illuminated when removing the USB in order to view
data.
2.2.
Documentation
During the test, both the bump test records and the event log of each unit will be downloaded to the
onboard USB memory. For more detailed information please refer to the
manual.
2.3.
Hibernation
C-Test has the ability to hibernate the Crowcon Clip+ monitor. To use this feature, power on the C-Test,
place the monitor in the bay, press and hold down the bump test button and calibration button at the
same time for 2 seconds. All lights will turn orange and the power light will blink. Once complete, the C-
Test lights will turn green for a successful hibernation and red for a failure. Please do not remove the
detectors until the power light turns green.
Note; during this process the event logs of the monitor are automatically transmitted to the USB in the
C-Test. Hibernating will erase the event log on the monitor.
3. Troubleshooting
If a monitor has failed a bump test make sure to check the following:
Check to see if any excessive dirt/debris has entered into either the sensor cavity or audible
alarm cavity. If so make sure to clear these before testing
Check to see if any excessive dirt/debris has accumulated over the small IR communication
window located on the top of the monitor parallel to the top light bar
Check the calibration gas level. 58L bottles are "full" at 500 PSI: 116L at 1000 PSI
Make sure the C-Test station is not placed directly in very bright light. Very bright light can
interfere with communication of the detector
Re-Try the unit. If a monitor has failed three (3) times please contact Crowcon or your local
service agent
Issue 1 - June 2012
GREEN
light; a failed test will result in a
(GREEN
or RED) you can then remove the
Page 4
RED
light.
Advanced
section of this