Warnings
Remember to wear safety glasses.
Leak test all fittings when using hazardous or flammable gases.
Do not use above 14 bar (200psi).
Do not open filter cartridge, even after use!
Points on Manifold block are sharp.
Special precautions might be required when using hydrogen. Consult local
regulations and your company's safety procedures.
The Advanced Filter System is intended to remove 50ppm or less of
oxygen, water, or hydrocarbons (C4 and above) from helium and other inert
gases, nitrogen or hydrogen at flow rates of 1L/minute or less.
Do not use to remove more than 50ppm of oxygen! Rapid heating of the
oxygen adsorbent will occur – and very high temperatures can result.
Limits:
Temperature
Pressure
Gas Supply
Flow Rate
CRS Advanced Filter System
15ºC to 35ºC
0.75 to 14 bar
< 50ppm O
, H
O and Hydrocarbons
2
2
< 1L/min