Connecting the Manifold
NOTE: Do NOT remove the black plastic covers from the check valves until manifold installation
is complete!
1.
The manifold block must be connected to the gas lines. The compression fittings on the block fit
1/8" stainless tubing. If you need to connect the filter to lines of another diameter, adapters will
need to be installed first.
2.
Select the bracket, bench-mount (L-shaped) or wall-mount (flat), which you intend to use. Use the
bracket first to mark the holes on the bench or wall where the manifold will be mounted.
The bench-mount bracket can also be attached without screws by using the supplied double sided
tape.
3.
Drill the holes to mount the bracket, either to fit the anchors provided or to fit the screws without
anchors, depending on the requirements. In the case of the Bench Mount I configuration (see
pictures below) the bracket should be screwed into position at this point.
4.
Insert the long hex-head screw into the back of the manifold if it is not already in position. Fasten
one of the brackets to the manifold, trapping the long screw.
5.
Fasten the bracket to the bench or wall in the correct position.
6.
Thread the nuts, back ferrules and front ferrules onto the tubing. Insert the tubing into the
manifold fittings, making sure that it reaches the bottom of the fitting.
7.
Tighten the compression fittings. Compression fittings normally require about ¾ turn past the
finger-tight position for a good seal. Do not overtighten or the fittings may be damaged.
8.
Remove the protective plastic cups from the manifold check valves. Be careful, the ends of the
check valves are sharp!
Inlet
6
Outlet
CRS Advanced Filter System