Direct Healthcare Group 56-303/75-B/0-PU Podręcznik użytkownika - Strona 6

Przeglądaj online lub pobierz pdf Podręcznik użytkownika dla Pomoc w poruszaniu się Direct Healthcare Group 56-303/75-B/0-PU. Direct Healthcare Group 56-303/75-B/0-PU 13 stron.

Direct Healthcare Group 56-303/75-B/0-PU Podręcznik użytkownika

10. Maintanance

Recommended at least once per year.
Check that
General
The Walker feels firm/does not feel loose.
The Walker does not rattle during manoeuvres.
The Walker is level and all of its castors are in contact with
the floor.
The Walker is not dirty.
Arm rests
Arm rests are intact and clean.
Arm rest widening is functional.
Handle
Grips are not damaged/dirty.
Handle adjustment is functional.
Battery
Check that the charging cable is connected to the control
box.
Check that the remote control is connected to
the control box.
Connect or fit new remote control as required.
Check that the battery, actuators and control box
are not loose.
Height/width adjustment
Raising and lowering are functional.
The Walker feels stable at maximum settings.
There is no looseness between the vertical frame and
bottom frame.
Height adjustment clamps lock.
The electric actuator must not feel loose.
Height adjustment via remote control is functional. The
electric actuator must run smoothly at a constant speed.
Frame parts
There is no mechanical damage.
There are no scratches.
End plugs/lower frame fitted.
Castors/brakes
Castors roll easily/tread not damaged.
The castors are tightly fastened to the lower frame.
Castor brakes functional on all castors.
Handbrake functional.
6
BUR E D O UB L E 2 .0
Action
Clean with lukewarm soap solution or alcohol-based
cleaning agent (no petroleum products).
Clean; replace if damaged.
Replace lock knob.
Clean with lukewarm soap solution (no petroleum
products).
Replace handle grips.
Fit lock knobs/plates as required.
Fit new cable; this must always be connected to the
control box (IP65 classification).
Connect; replace with new remote control as required.
Refer to the wiring diagram in the Technical manual.
Tighten; replace with new fasteners as required.
Fit new guide sleeves (in frame) or end plugs
(in chrome tubes) as required.
The technical manual shows the design and which
fasteners must be checked. Contact the distributor if the
parts cannot be tightened.
Replace clamps.
Tighten the attachment fitting concerned; replace bolts/
lock nuts as necessary.
Check that the battery/remote control/actuator are
connected in accordance with the wiring diagram.
Charge the battery. Contact customer services.
If damage is present, contact GATE's customer services.
Touch up as necessary.
Fit new end plugs.
Clean or replace castors (NOTE: Castors are always fixed
to the lower frame with thread locker or lock nuts. The
castors are sealed and we do not recommend their
dismantling; instead, replace the whole castor assembly).
Tighten the castor bolt and fit locking nut or use thread
locker (depends on model).
Replace castor assembly.
Adjust the brake or fit new castors.
DIR ECTHEALTH CA RE GR OUP. C OM