7. Control Box Wiring
Remove the control box cover to make the connections as described
below. Use a screwdriver or similar; refer to the special labels on the
control box.
NOTE! Make sure that plugs are connected to the correct sockets
otherwise the actuators can be damaged.
1. Battery terminal socket.
2. Plug socket. NOTE: The plug must ALWAYS be connected to the
control box for IP protection to be valid.
3. Remote control socket.
4. Socket for the raise/lower actuator.
8. Battery Charging
Battery charging must take place when the Walker is not in use (the battery
can never be overcharged). Charging is carried out by inserting the plug in
a wall socket.
The battery should be charged at least 24 hours before first time use.
9. LED on Control Box and Battery
Control box – green light when power is available from battery/mains
socket.
Battery:
•
The LED flashes green during charging.
•
ON/OFF: To use the product, press ON.
•
The green lamp will begin to flash.
•
Use the OFF button (hold down for 5 s) to switch off the battery if the
product will not be used for an extended period.
•
The LED will shows a constant green light when the battery is fully
charged.
•
When the battery is in normal state/in operation, the LED lashes green
at REGULAR intervals, namely 0.5 s ON, 4 s OFF.
•
When the battery requires charging (remaining charge <25%) the LED
will flash orange/red at REGULAR intervals, namely 0.5 s ON, 4 s OFF.
NOTE: The battery has an in-built device that automatically switches it off
when remaining charge is <20%. The LED is then extinguished. The battery
must now be fully charged before the electronics can be used again.