Direct Supply Panacea 3500 Podręcznik konserwacji - Strona 5

Przeglądaj online lub pobierz pdf Podręcznik konserwacji dla Sprzęt medyczny Direct Supply Panacea 3500. Direct Supply Panacea 3500 8 stron. Soft touch siderails for beds
Również dla Direct Supply Panacea 3500: Instrukcja obsługi (12 strony)

Direct Supply Panacea 3500 Podręcznik konserwacji
Step 10:
Inspect all bolts and fasteners. Do
not over tighten bolts at pivot points.
Step 11:
Check all cords for cuts or
other damage.
Step 12:
Make sure all plugs are
fully seated.
15-Point Bed Inspection
Step 13:
Make sure all cords are free of
moving parts.
Step 14:
Visually examine all components for
damage or excessive wear, cracked or missing
brake pads. Make sure retention clips and pins
are secure.
Step 15:
Check that brake pads are intact
and offer sufficient resistance to hold the bed
securely with a horizontal force of 65 pounds.
If this test fails, it is recommended that the
brake pads be replaced.
1-800-634-7328
DirectSupply.com
Continued
5