V.PHOTOMICROGRAPHY
Prepare the following equipments in addition to
the LABOPHOT-POL microscope main body.
*
Nikon Microflex
*
Trinocular eyepiece tube "TP"
*
CF PL Projection Lens
1. CF Pl Projection lenses
The combined use of the CF P objectives and
CF PL Projection lenses is essential.
For the same total magnification,
select a com-
bination of the highest possible objective power
and lowest possible projection
lens power to
achieve
the
utmost
image
definition
and
contrast.
2. Illumination
1) Checkingthe illumination
Unevenness in the illumination
will show
up
more
conspicuously
in photomicro-
graphy than in observation. Consequently,
before taking a photograph,
recheck the
positioning
and centering of the lamp and
the correct adjustment of the condenser.
2) Selection of voltage and filter
The color temperatu re of the Iight source
varies with the voltage being used. There-
fore,
in
color
photomicrography,
the
selection of voltage and filter
is essential
(for the result to be obtained).
In color photomicrography,
set the bright-
ness control
dial to 5.5, and use NCB10
filter.
Depending
upon the make of the film,
different
color renditions may result. It is
recommended that in addition to the NCB
10 filter
a color compensation
filter (CC
filter),
available from the film
manufac-
turer, be used.
3. Shutter Speed
Desirable shutter speeds for least vibration are
1/4 ~ 1/15 sec. Adjustment of the image bright-
ness for
color
photomicrography
should be
made by means of the NO filters.
Some specimens require, on account of their
15
insufficient
brightness, longer exposure times,
and consequently
poor color
reproducibility
owing to the "Reprocity
Law Failure" of film
may result. So, when taking picture of such
specimens, it is recommended to use the N ikon
Polarizing Microscope OPTIPHOT-POL.
4. Manipulation of Field and Aperture
Diaphragm
In photomicrography,
the adjustment
of the
field diaphragm is important for the purpose of
limiting extraneous light which causes flare in
the microscope
image. Stop down
the dia-
phragm so as to get an illuminated area slightly
larger than that of the picture field. By adjust-
ing the apertu re diaphragm, a change of depth
of focus, contrast and resolution
of image is
attainable. Select a size suited to the purpose.
Generally speaking, the aperture diaphragm, is
properly
stopped down to 70 ~ 80% of the
aperture of the objective being used.
5. FOC;USing
Focusing for photomicrography
can be done
with
the observation
tube of the trinocular
eyepiece tube "TP"
or by using the Microflex
finder.
1) For focusing with the Microflex finder
Refer to the Instruction
Manual for the
Nikon Microflex.
2) Focusingwith the observation tube
For focusing with the observation tube, use
the eyepiece incorporating the photo mask.
Before proceeding to focusing, the bino-
cular diopter adjustment should have been
finished.
(1) Insert the eyepiece with photo mask into
the eyepiece sleeve on the side of the user's
dominant
eye, and the viewing eyepiece
into the other side sleeve.
Turning the diopter ring, bring the double
cross lines in the mask eyepiece into sharp
focus, and then, turning
the coarse-fine
focus knob, focus the specimen image onto
the focused surface at the center of the