Calculation of Magnetite from MF500M readings
The magnetite meter measures the force on a powerful rare earth
magnet caused by the ferritic component of the magnetite. It is a
prerequisite of this technique that the containing pipe is stainless
steel which is not magnetic. In making the measurement the
instrument has a sensor that has some compensation for the
probe orientation in respect to gravitational forces.
Clearly the magnetite could be in any amount at any point in the
pipe, but in reality there are only really 2 likely scenarios. Firstly
the pipe is fully filled and in normal use there is continuously
flowing water filling the tube. In this case the magnetite is likely
to be evenly deposited on the inside surface. (Note this is not
guaranteed and could be affected by turbulent flow or surface
finish attributes). The second scenario is that the magnetite has
accumulated on one side of the pipe. This can be due to a
horizontal pipe that is not fully filled, or accumulation of flakes of
magnetite building up under gravity. The measurement rote
attempts to identify which type of deposit is in place and then
estimate the deposit thickness from the measurement.
Before starting the probe should be placed in a position at least
1m from any magnetic material and the instrument should be
zeroed. This removes any constant offset in the measurement
system to be removed from all subsequent measurements.