Chapter 3: Trouble shooting

FAULT
PROBLEM
CODE
Instruments can't be power
E01
on
Instrument
E02
movement
E03
LCD
display
characters
E04
Instrument speed is not
accurate
Self-test, TE1 Ref,TE2 Ref,
TE3 Ref--error
CAUSE
The
power
unplugged
The power switch put off
The fuse is broken
Power board is failure
Main board is failure
The display screen is
broken
No
setting
temperature
The drive board is failure
doesn't
24V
power
failure
Motor is failure
Check whether shrapnel
is broken
Belt is loosen or broken
garbled
Display board is failure
Photosensor is failure
The
position
photosensor goes failure
8
line
is
Check whether the power line is
unplugged, an power on it again
Put on the power switch
Replace the fuse
Replace power board
Replace mainbaord
Replace display screen
target
Set a target temperature, and
the temperature indicator is on.
Replace the drive board, please
reference chapter 2.5
board
is
Replace
component
Replace motor
Replace shrapnel.
Replace belt
Replace display board
Replace heated lid
of
Adjust
photosensor
SOLUTION
power
board
the
position
of