Foot electrode calibration
Place a measuring electrode on the foot
1.
electrode you want to measure and connect it
to a resistance decade
Connect the calibration test point of the tester
2.
to the resistance decade
To calibrate, adjust the knobs on the resistance
3.
decade and start testing by touching both parts
of the touch button simultaneously
Wrist strap/groundable ESD smock connection calibration
Connect the wrist strap/groundable ESD smock
1.
connection to a resistance decade
Connect the calibration test point of the tester
2.
to the resistance decade