Chapter 2 – Installation
2.1 Mounting
WARNING:
the units when opening if they are not permanently mounted. Care must be
taken when assembling the two units together before mounting.
Carefully feed wiring and tubing bundles
from the bottom of CL8 into
corresponding holes on top of the
pyrolyzer. Make sure the grommets are
attached over the bundles inside the
CL8.
Place CL8 unit on top of pyrolyzer. Align
units properly and make sure wires &
tubing are not pinched. Securely tighten
screws from inside pyrolyzer.
Press the Grommets into the hole to
cover the metal edges of both the CL8
and the pyrolyzer. Check the Grommet
from above and below to assure they
are properly seated
DC-ITD-CL4-CL8MANPYRO_A
REV 8/25/2022
The doors on the CL8 unit and the pyrolyzer are heavy and can tip
Wiring & Tubing
bundles exit holes
Pyrolyzer unit insert
and tighten four (4)
8/32" screws at the
top and tighten to
secure unit to CL8
using pre-tapped
Grommets seated
Protective Earth (PE)
on bottom of
CL8 unit.
From inside the
holes.
properly.
Reattach the
ground wire inside
the CL8 cabinet
5
CL4 & CL8 Manual