Cane CRONO Super PID Skrócona instrukcja obsługi

Przeglądaj online lub pobierz pdf Skrócona instrukcja obsługi dla Sprzęt medyczny Cane CRONO Super PID. Cane CRONO Super PID 2 stron. Ambulatory infusion pump

CRONO
ambulatory Infusion pump
A B B R E V I AT E D U S E R G U I D E
Keep buttons depressed for a few seconds to activate commands.
When the pump is locked the syringe symbol blinks.
Battery
compartment
ButtonS
p
1st depressing
p
2nd depressing
p
3rd depressing
-
+
/
.
-
p
and
depress simultaneously
-
depressed for 8 seconds
+
ButtonS
p
.
-
+
/
-
+
and
depress simultaneously
Super pID
Battery InSertIon phaSe
• Pump safety test
• Pusher self-adjust
• Pump switches OFF
programmIng
Conditions for programming:
- Pump in OFF;
- Start of a new infusion;
- Pump unlocked;
• Syringe size setting (10 / 20 ml)
• End of infusion acoustic signal setting: OFF / ON
• Partial volume setting:
• Decrease/Increase above parameters
puSher reverSIng / numBer oF InFuSIonS DeLIvereD
• Pusher reversing to the infusion starting position
• Number of infusions delivered
• Pump switches ON
• Pump unlocked: 0.2 ml each depressing
• Pump locked: maximum 10 boluses or 2 minutes for priming
programmIng
• Delivery time setting (pump unlocked):
- from 1 h to 99 h with 15 min steps
- from 15 min to 1 h with 5 min steps (10 ml syringe)
- from 30 min to 1 h with 5 min steps (20 ml syringe)
• Pump switches OFF
• End of infusion
• Pusher automatic reversing
• Automatic switch OFF
1-20 ml in steps of 1 ml
pump on
prIme
pump oFF
Screen
10
20
20
20
Screen
10
20
20
20
20
20
20
10
20
20
10
20
Screen
20
20
10
20
20
Screen
20
20
20
20