TROUBLE SHOOTING
PROBLEM: Backrest does not function properly.
SOLUTION: If the headrest does not remain flat in
the down position, loosen nut "A" and
tighten nut "B."
If headrest does not move at all, loosen
nut "B" and tighten nut "A."
After proper function has been achieved,
tighten both nuts.
PROBLEM: Table height adjustment will not work properly.
SOLUTION: Make sure the table is connected to a working electrical outlet. Inspect all power cords
for cuts or breaks that would interfere with the electrical connection. Check to make certain foot
pedal cord is properly and securely plugged into motor. If after checking all these areas, the table will
not operate, or operates improperly, call a technician, or call Clinton Industries at 800-441-9131 for
assistance.
REMOVAL OF THE MOTOR FOR SERVICE
The motor assembly is located at the foot-end of the table on the base. Next, detach the motor
assembly. If possible, raise the table to its highest position. Raising the table will make it easier to
remove the motor mount bolts. Using help, lower the table onto its side onto a protective surface
such as a blanket or other material that will not harm the upholstered top. Remove the foot pedal
controller by unplugging the DIN plug on the side of the motor. Carefully remove the upper and
lower motor mount bolts to free the motor assembly.
(continued)
Nut A
Upper Motor Mount
Motor Assembly
Lower Motor Mount
Foot Pedal DIN Plug
Nut B