Etac Molift Nomad Instrukcja obsługi - Strona 8
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Sprzęt medyczny Etac Molift Nomad. Etac Molift Nomad 18 stron.

Operation
Functions
Emergency lowering
Emergency
stop
Down
Service light
Contact hand control
Hand control
Operating the lifter
Connect the hand control cable to the plug on the
back side of the lifter. The hand control has two
buttons which gives the operator excellent control
of the movement of the lifter up and down. Two
buttons with the same function are placed on the
right front side of the lifter.
Emergency stop and emergency lowering
All Molift lifters are equipped with emergency stop
and lowering. The emergency stop button will cut
the power to the motor, and is situated on the left
front side of the chassis. Turn the button clockwise
to reset. Charging and emergency lowering does
not work if the emergency stop is activated.
It is possible to perform an emergency lowering in
case of an electronic malfunction or uncontrolled
stop. There is an emergency lowering button on
the left front side of the chassis with a yellow seal.
Move the lifter to a suitable location to lower the
user.
• Puncture the seal and press the emergency
lowering button. The lifter will make a smoothly
movement downwards.
Page 8 of 20
• If the lifter don't move, charge the battery, and
try the emergency lowering button again. Do
not run the lifter when connected to the charger!
• If none of the actions above works, the user
must be lowered manually. Call for assistance.
Up
Battery light
Battery care
The batteries of the Molift NOMAD is a 26,4 V cell
unit that is placed inside the chassis.
The lifter has a battery light on the hand control
which illuminates when the battery voltage drops
below a certain level. If there is no hand control
connected the battery light on the lifter will illu-
minate when battery level is low. There will also be
an audible alarm, which indicates that the battery
needs charging. There is still enough capacity to
lower the patient.
BM15101 Molift NOMAD English - Rev. M 23-08-2013
If the Molift Nomad is operated by the user
alone, the Emergency Stop and Emergency
lowering button is difficult to operate. Make
sure to have a mobile telephone or other com-
munication device available.
If the reason for failure causing an emergency
lowering is unknown or if it is a result of failed
component, contact your local service repre-
sentative/dealer or Etac Supply Gjøvik.
If the battery is discharged during a lift, it is
always enough power to lower the user down
again.