If the lift is not working properly following assembly (page 4) and the pre-use checklist (page 3), the
following troubleshooting table may help you to pinpoint the fault:
Symptom
The lift column is unstable.
The lift does not roll straight on level
flooring.
The lift only goes up or down/the legs
only go in or out.
The lift runs by itself/the leg adjust-
ment works by itself .
The lifting arm does not move.
The battery does not charge.
If the troubleshooting does not help to resolve your current problem, please contact the relevant service
personnel; your local representative or Etac AS may also be able to refer you to someone to help you with
the problem.
BM08101 Molift MOVER 180 English - Rev C 19.12.2013

Troubleshooting

Possible cause/solution
Check the lift column's holder and both allen screws as described in
the assembly instructions on page 4.
Clean the wheels
Contact your local service representative.
Try a different hand control if you have one; you may need to replace
the faulty hand control.
Dismantle column to check contact point between chassis and column
Contact your local service representative.
Clean the hand control contact with spirit to remove any grease.
Replace the hand control/contact your local service representative.
Contact your local service representative.
Make sure that the emergency stop is not activated (turn clockwise to
reset the emergency stop
Conduct an inspection/change to a different battery or charge the bat-
tery
The lift's electronics have overheated; allow the lift to cool down.
Make sure that the load weighs no more than 182 kg.
Make sure that the hand control contact has not come loose.
Try a different hand control if you have one; you may need to replace
the faulty hand control.
Contact your local service representative; you may need to use the
manual emergency lowering function to lower the patient.
There may be a fault with the battery; try a different battery or replace
the battery.
There may be a fault with the battery charger; try a different charger/
contact your local service representative.
Page 13 of 16