1. Lumaquant software offered by Etaluma (see Lumaquant Startup Guide for compiling images into
video,
2. Your own compiling software such as ImageJ, Fiji, MSMoviemaker, or iMovie, or
3. Use of two freeware software programs together, i.e., AviSynth and MeGUI (instructions described in
Appendix C). If using this option, be sure to read the instructions first because the two programs must
be downloaded before image capture.
I.

Time-Lapse

1. To configure a time-lapse assay, the Protocol feature is utilized.
Click the "P" icon in the Left Tool Bar (fourth from the top) to
open its dialog box containing three tabs. Enter information and
values as you move between the tabs. Be sure to click Save
before leaving a Protocol.
2. In the first tab called Information, click the Load Existing button
to bring up a previously saved Protocol (select desired Protocol
and click OK). Click the Clear Existing button if values have been
entered on any of the tabs and not saved. A Save Protocol File
dialog box will open asking if you want to save the current
Protocol. If Yes, click OK. If not previously saved, enter the new
desired Protocol name in the Filename field or click on an
existing File name to overwrite. If the latter, a File Overwrite
Alert will open; enter Yes or No. You are now ready to create a
the Information tab, if you enter User Notes, they will become part of the new Protocol. You can also
enter User Notes for a previously saved Protocol in this tab.
3. Click on the Acquisition tab and check Time-Lapse. Enter the
Interval and Duration information. Before initiating the time-
lapse run, review Computer Configuration in the Help Section
and change your computer power and Windows update settings
so they cannot automatically turn on and interfere with
scheduled image captures. Of particular importance to turn off
are Sleep Mode and Auto Restart. (After time-lapse is
completed, be sure to change the settings back.)
4. Click on the Image tab to select the channels to be used in the
time-lapse and adjust their settings (LS460 will show only the
Brightfield settings). If settings were previously determined in
Manual Image (Section II.D above), transfer the values into the
current Protocol by clicking the Use Manual Settings button. The
desired values can also be entered manually. Note: This dialog box
cannot be used to optimize the image settings because it is not
interactive with the Live Image Window.
If optimization of any of the Channel settings has not been done or is still needed, click on the Manual
Image icon on the Left Toolbar (top icon) to open its dialog box and fine tune the settings with a
2017.11.20
LS560/460 Startup Guide
new Protocol. On