Stone model and non-adjustable articulator scan : After placing the stone model and
❸
non-adjustable articulator on the scanner you can proceed with scanning. The camera
moves up and down to scan the occlusal relationship.
Adjustable articulator : After removing all base plates, place the entire adjustable
❹
articulator on the scanner. By removing the articulator incisor pin, you can scan for a
better occlusal relationship.
34
Scanning Instructions
Sort scan data
9
Scan data alignment occurs automatically in the background while scanning is in
❶
progress.
You can check the sorted data when the scanning process for one arch is completed.
❷
For example, in a scanning process that proceeds in the order of base, abutment and
❸
wax-up, the sorted data can be checked after the wax-up scan is completed.
Sorting may fail due to dissimilar data or errors. In this case, you can manually sort the
❹
data by clicking the Matching button.
Sort the data by clicking Matching using 3 points in common areas. Sorting is possible
❺
with just 1 or 2 points. When sorting data, dots must be set equaliy in common areas
of left and right scan data.
Scanning Instructions
35