ASG DTT-L Series Podręcznik użytkownika - Strona 8

Przeglądaj online lub pobierz pdf Podręcznik użytkownika dla Sprzęt testowy ASG DTT-L Series. ASG DTT-L Series 15 stron. Digital torque testers

ASG, Division of Jergens, Inc.
S. Waterloo Road | Cleveland, OH
15700
DTT-L Series Digital Torque Testers
User Manual

Battery Indicator

If the battery level is less than 4.6 V, the "Battery Empty" message will be displayed and the tester
will power down automatically. NOTE: Only use the 9V adapter/charger supplied.
Battery

Basic Function

Battery Level > 5.00V
TRACK
4.99V > Battery Level > 4.8V
0.003
4.79V > Battery Level > 4.7V
4.69V > Battery Level > 4.65V
TRACK
Battery Level < 4.65V
0.003
Basic Function
Clockwise torque is displayed on the DTT-L and recognized by the symbol "CW." Counter-clockwise
torque is displayed on the DTT-L and recognized by the symbol "CCW."

Display of Clockwise/Counter-Clockwise

Load indicator bar
for clockwise
Load indicator
bar for counter-
clockwise
A load indicator bar alerts the operator as to how much torque load is being applied to the
transducer as related to its full scale rating.
| Phone: (
44110-3898
888) 486-6163
005
005
F-PEAK
0.003
kgf.cm
kgf.cm
CW
CCW
005
005
PEAK
0.003
kgf.cm
kgf.cm
CCW
CCW
Clockwise symbol
Counter-clockwise symbol
| Fax: (
| Email: [email protected] | www.asg-jergens.com
216) 481-4519
Charging symbol
Page 8 of 15