Installation and Maintenance Instructions
for 50 mm HPS High Purity Gauge
WARNING
Do not use this pressure gauge with fluids that may cor-
rode the instrument's wetted parts.
If corrosive fluid flows through this pressure gauge, the
pressure element (Bourdon tube) may be damaged or
burst and it flows out, resulting in injury or damages to the
peripheral device and equipment.
5.6
Do not modify the pressure gauge.
WARNING
Do not attempt to modify the product itself or provide it
with additional functions.
5.7
Do not apply glue to the blow-out disk, tear a hole, inter-
space, or make any other alterations to the instrument. Duly
note, if ignored, the unit will fail to function as designed.
Failure to relieve pressure properly, this could result in quite
.
dangerous accidents
6. INSPECTION AND MAINTENANCE
6.1
Check the gauge accuracy regularly to ensure correct indi-
cation; this is recommended to be once or more per year.
6.2
Pressure indication error of approximately 1 graduation is
repairable, yet the instrument must be replaced when the
error is identified as substantial.
7. TROUBLESHOOTING
Refer to the accompanying table and implement appropriate
countermeasures. Contact Ashcroft in the event the issue
persists.
© Ashcroft Inc. 2022, 250 East Main Street, Stratford, CT 06614-5145, USA, Tel: 203-378-8281, Fax: 203-385-0357, www.ashcroft.com
All sales subject to standard terms and conditions of sale. 50_HPS_Gauge_I&M008-10305_RevA_02-10-22