Ashcroft DG25 Instrukcja instalacji i obsługi - Strona 9

Przeglądaj online lub pobierz pdf Instrukcja instalacji i obsługi dla Przyrządy pomiarowe Ashcroft DG25. Ashcroft DG25 16 stron. Digital pressure gauge

Ashcroft DG25 Instrukcja instalacji i obsługi
Maximum / Minimum:
Displays maximum / minimum pressure values;
this is initiated upon powering the unit or since
the values were cleared.
1. When in menu mode press
to scroll to 'Max' (maximum display);
this is indicated in small font to the left
of the display's unit of measure.
2. To clear oth Min and Max values, press
©
and hold
ZERO
3. Release of this key will leave you in
Menu mode. Use
continue scrolling through the menu
options, OR hold,
to measurement mode.
Note: Clearing Minimum / Maximum
values will reflect a blank display except
for the associated Min. / Max. icon and
battery indicator.
Note: The following actions will also
clear min/max values. Power off, zero
gauge function, tare function, field cali-
bration function, units programming,
update rate, or reset.
©
©
,
.
©
©
,
to
©
to return
M ENU
Timer:
Controls how long the gauge will remain pow-
ered ON once the Power key
1. When in menu mode press
to scroll until "TIMER" is displayed.
©
k
2. Press
3. Then, press
through values. "NONE" designates that
the gauge will remain ON until the
©
k
power key
time. Any other value (1 min [default] /
5 min / 20 min), other than "NONE", will
designate the duration of time this func-
tion will be in effect.
©
k
4. Press
timer icon will be shown on the display
and the unit will display "DONE" then re-
turn to Measurement Mode.
Feature: Timer icon will flash 10 seconds
prior to gauge shut down.
9
©
k
is pressed.
©
©
,
.
©
©
,
to scroll
is pressed a second
to select desired value;