Dolomite Mitos Instrukcja programowania - Strona 8

Przeglądaj online lub pobierz pdf Instrukcja programowania dla Pompa wodna Dolomite Mitos. Dolomite Mitos 15 stron. Automated product collector
Również dla Dolomite Mitos: Skrócona instrukcja obsługi (3 strony)

Dolomite Mitos Instrukcja programowania
The Dolomite Centre Ltd.
Command Description
Response "#A0"
If pump does not receive any command, e.g. a status query "s", for 30 seconds it will
automatically return to manual mode.
Your application sends this to move the pump
C
out of ERROR state and back into IDLE. C will
also stop control and force the pump back to
IDLE.
You must explicitly clear all error conditions.
This ensures your application has noted the error
and corrected it. The ERROR state will not clear
unless any physical error condition has been
corrected.
e
Returns text string of the last error message.
This may a P-Pump specific error or an error
from the underlying micro controller board.
Note, this command will always return the last
error message no matter how old. The message
is only valid when the pump is in the ERROR
state. See section 6.4 below, Error codes, below
Returns two strings: date and error information
separated by <LF>.
L
"Lpump Z" sets display name on the pump to
"pump Z". The label is limited to a max of 8
characters - alphanumeric plus underscores.
Name will be upper case only.
l
Lower case L.
Returns current label on the pump,
e.g. "#lpump Z"
Returns the serial number for the pump,
n
e.g. "#n160295"
Send "Tx" where x is UTC time in seconds since
T
1/1/1970
t
Responds "#txxxxxx" where xxxxxx is UTC time in
seconds since 1/1/1970
Replies with "#sErr,Sp,Src,Pc,Ps,Pt,Qc,Qt,Ft", see
s
section 6.3, Status command response, below
for details.
P
"P1000" sets target pressure to 1000 mbar and
starts control. Replies "#P0" and starts pressure
control if in remote control mode (see A1).
Setting target pressure to 0 (sending the "P0"
command) moves the pump to the IDLE state. It
means the pump stops controlling pressure and
opens the vent valve.
3600051 - P-Pump API - programming instructions - v 1.3
Pump State
Control Mode
(before after)
(before after)
ERROR
IDLE
(ANY
IDLE)
ANY
ANY
[response
only
guaranteed
valid if in
ERROR state]
IDLE
Remote
ANY
ANY
ANY
ANY
IDLE
ANY
ANY
ANY
ANY
or
ANY
ERROR IDLE
IDLE
CONTROL,
CONTROL
REMOTE
CONTROL
or CONTROL
IDLE
Page 8 of 15