ASI 1045 Instrukcja obsługi i serwisowania - Strona 15
Przeglądaj online lub pobierz pdf Instrukcja obsługi i serwisowania dla Sprzęt dentystyczny ASI 1045. ASI 1045 20 stron. Dental operative unit classic assistant’s cart
c
laSSic
Caution: Do not use the DentaPure DP365 Cartridge for longer than its rated one-year period. The cartridge will no
longer provide disinfection for the system and may plug the unit and obstruct the water flow.
Caution: Do not use the DentaPure DP365 if using a shock treatment to disinfect lines as it will damage the cartridge.
The cartridge must be removed and a standard tubing used in the bottle for performing shock disinfection treatment of
the water system.
Caution: Do not use bleach to disinfect the water system. Due to its caustic nature, it can corrode certain internal
parts. Any damage is excluded from warranty coverage.
Caution: The user assumes all responsibility for maintaining and disinfection of the water system, and for periodically
having water sample tests performed to ensure that disinfection is satisfactory.
There are many other established methods for disinfecting a closed water system that consist of liquids, powders and tablets.
However, the chemicals used in these cleaners may damage the materials of the diaphragms used in the control block and
other valves in the delivery system. Further, tablets may not dissolve completely and may cause blockage of the water system.
Any damage to the system as a result of the use of these alternate cleaners is excluded from warranty coverage.
As an alternate, the bottles may be filled with a solution of 10 parts water and 1 part mouthwash that contains alcohol.
For more information on this protocol, please see abstract entitled "A Chemical Treatment Regimen to Reduce Bacterial
Contamination in Dental Water Lines" in the May 1997 issue of JADA. If using another type of disinfectant, follow the
instructions provided with your disinfectant for effective control of the growth of biofilms in the water line.
3.3.4 Syringe Tips
For reusable tips follow the manufacturer's suggested guidelines for tip cleaning.
3.3.5 Sterilization
The System and the Attachments offered with it (Air/water syringe, Saliva ejector and High volume evacuator) are not provided
sterile. The air/water syringe and suction handpieces are designed for use with disposable tips. If using reusable tips supplied
by others, they must be properly sterilized before each use according to their manufacturer's instructions. In addition to the
disinfection of these parts, a disposable barrier sleeve must be used to cover the air/water syringe, saliva ejector and high
volume evacuator (HVE). The disposable barrier sleeve must be discarded and replaced between each patient.
Follow all manufacturer's instructions for sterilization of any optional accessories and dental handpieces made by others.
3.4
Preventative Maintenance Schedule
See complete schedule in appendix, section A4
Rev.A
ASI - The leader in Advanced Systems Integration www.asimedical.net
a
1044/1045/2044/2045
dvanced
m
anual
15