Step
Action
7
Slip the Pop-Top clamp ring over the cap. Slide the clamp ring
around the "fingers" of the cap until it is firmly seated.
Autoclaving
Pop-Top holders, both unloaded or loaded with Whatman™
membranes, can be repeatedly autoclaved at 121°C (250°F) for 15
minutes at 0.1 MPa (1.0 bar, 15 psi).
When autoclaving loaded Pop-Top holders, loosen the clamp ring
about half way from the cap fingers. Reseat clamp ring after auto-
claving.
Operating precautions
•
Make sure that flow direction is from the inlet in the cap (female
Luer slip fitting) to the outlet in the base (male Luer slip fitting).
NOTICE
Flow in the opposite, unsupported direction
could rupture the membrane.
•
Make sure the clamp ring is flushed with the top of the cap
"fingers" before applying pressure so that maximum leak tight-
ness is achieved.
CAUTION
Never operate the holder above 0.34 MPa
(3.4 bar, 50 psi) at 38°C (100°F).
CAUTION
Never use the holder with fluid temperatures
over 140°C (285°F).
Give feedback on this document
Visit
cytiva.com/techdocfeedback
scan the QR code.
cytiva.com
Cytiva and the Drop logo are trademarks of Life Sciences IP Holdings Corp. or an
affiliate doing business as Cytiva.
Whatman is a trademark of Global Life Sciences Solutions USA LLC or an affiliate
doing business as Cytiva.
Any other third-party trademarks are the property of their respective owners.
© 2023 Cytiva
For local office contact information, visit cytiva.com/contact
29690340 AB V:8 03/2023
or
Disassembly
To disassemble the product, follow the instructions below.
Step
Action
1
Slide the clamp ring away from the "fingers" of the cap.
2
Grip the cap in one hand using the thumb and forefinger. With
the other hand, apply pressure (torque) to the male Luer exten-
sion with the forefinger and thumb until the base snaps loose.
Cleaning
To clean the holder, follow the steps below.
Step
Action
1
Completely disassemble the holder.
2
Wash all parts with a suitable solvent (isopropyl alcohol, for
example). If scrubbing is required, use a soft bristle brush.
NOTICE
Do not use acetone as this can harm the
holder.
3
Rinse the disassembled holder in clean water and air dry.
4
Store the disassembled holder in a plastic bag or other clean
receptacle.
Note:
If the holder is pre-loaded and autoclaved, do not store for an
extended period above normal room temperature as, over
time, higher temperatures may cause the membrane to stick to
the o-rings and cause issues during disassembly.
Ordering information
Product code Description
420100
13 mm Pop-Top Filter Holder
Cytiva offers a range of membrane types and pore sizes for use with
the 13 mm Pop-Top Filter holder. For more information on these or
any other Cytiva products, visit cytiva.com.
Qty/Pk
10