Step
Action
3
Stop the flow and wait 15 minutes until the temperature is
constant in the range of 20-30℃.
4
Read the displayed conductivity value and compare it to the
theoretical value at the temperature of the calibration solution
from the graph below.
No further actions are necessary if the displayed value and the
theoretical value correspond. If not, proceed with the following steps:
5
Select System:Calibrate.
Result: The Calibration dialog box opens.
6
In the Monitor box, select CIT101_Cal or CIT102_Cal.
7
Type the theoretical conductivity value according to the graph
below in the Reference Value 1 box.
8
Click the Read Value 1 button.
Result: The measured value is displayed in the Measured
value box. The new constant value is updated and displayed in
the Cell constant box.
Click the Close button to exit the dialog box.
9
Fig 1. Conductivity of 1.00 M NaCl at 20-30℃
pH/C-902 Instruction 14003958 AB
Preparing and inserting the pH electrode
The tip of the pH electrode consists of a thin glass membrane.
protect it from breakage, contamination and drying out or the
electrode will be destroyed. Always store the electrode with the end
cover filled with a 1:1 mixture of pH 4 buffer and 1 MKO
in water only.
When it is not in use, e.g. during system CIP, service or storage of the
system, the pH electrode is replaced by a pH plug and the pH
electrode is placed in a holder that is attached to the frame adjacent
to the pH monitor
The instruction below describes how to insert the pH electrode:
Step
Action
1
Unscrew the plastic fastening nut from the pH cell and remove
the pH plug.
2
Remove the pH electrode from its holder and take off the
protective cap. rinse the electrode with pure water.
Note:
The pH plug may be placed in the pH electrode holder while the
electrode is in use
3
Remove the fastening nut from the pH plug, insert the
electrode into the electrode adaptor and slide the nut over the
adaptor.
4
Carefully insert the electrode in the flow cell. Tighten the nut by
hand to secure the electrode.
5
Connect the monitor cable to the electrode.
6
Connect the grounding wire to the pH cell.
Fig 2. Inserting pH electrode into the flow cell
Note: if the electrode is not fully inserted, the system will leak and a
dead volume will occur in the holder.
CAUTION
Handle the pH electrode with care
Note: When running chromatography using organic solvents it is
recommended that the pH electrode is removed and the
dummy electrode is inserted in its place, as organic solvents
will cause pH electrode degeneration.
. Do not store