EG&G ORTEC 414A Instrukcja obsługi i serwisowania - Strona 10
Przeglądaj online lub pobierz pdf Instrukcja obsługi i serwisowania dla Wzmacniacz EG&G ORTEC 414A. EG&G ORTEC 414A 15 stron. Fast coincidence
6. MAINTENANCE
I
ai. TESTING PERFORMANCE
The following paragraphs are Intended as aids in the
installation and checkout of the 414A. These instructions
present information on front panel controls, waveforms, at
test points, and output connectors. The following, or
equivalent, test equipment is needed:
ORTEC 419 Precision Pulse Generator
Tektronix Model 580 Oscilloscope
100S2 BNC terminators
Vacuum tube voltmeter
ORTEC 410 Linear Amplifier
Two ORTEC 420A Timing Single Channel Analyzers
Schematic and block diagram for 414A
Preliminary procedures consist of the following:
1. Visually check module for possible damage due to
shipment.
2. Connect ac power to ORTEC 401 /402.
3. Plug module into Bin and check for proper mechanical
alignment.
4. Switch on ac power and check the dc Power Supply
voltages at the test points on the 402 control panel.
Testing the
performance of the 414A
involves the
following:
1.
Feed the output of the 420A into coincidence Input
A. Terminate with lOOfi the RG-62/U connecting cable
feeding Input A.
2.
Set the In/Out control toggle switches as follows: A,
In; B, C, and D, Out.
3.
Set the Resolving Time control to 100.
4.
Observe the Fast Coincidence output pulse; it should
be a positive pulse, 6.5 ± 0.6 V and 0.5 ± 0.1 /isec wide.
Loading the 414A Output with 100J2 should reduce the
output to not less than 5 V and the width to not less than
0.5psec.
5.
Putting In/Out switch B or C to the In position should
inhibit the pulse on the 414A Output. Check that both B
and C inhibit. Connect the input into Input 8 and check
that both A and C inhibit. Connect the input into Input C
and check that both A and B inhibit.
6.
Connect the equipment as shown in Fig. 6.1; Keep all
coaxial cables terminated in their characteristic impedance.
7.
Ensure that a positive logic pulse is being received
from the output of each 420A. Set the output delay
control on the Calibrated 420A to 200 and adjust the
Reference 420A output pulse until the leading edges of the
pulses are approximately in coincidence.
414A
FAST COINCIDENCE
RESOLVING TIME
Referenc«
Out out out out
Calibrated
300673
PRECISION PULSE
GENERATOR
419
LINEAR
AMPLIFIER
410
TIMING
SINGLE CHANNEL
ANALYZER
420A
TIMING
SINGLE CHANNEL
ANALYZER
420A
Channel
Exterior
1
Trigger
OSCILLOSCOPE
Fig. 6.1. 414A Test and Calibrate Circuit.