Askle Sante Winncare Automorpho Plus Podręcznik użytkownika - Strona 15
Przeglądaj online lub pobierz pdf Podręcznik użytkownika dla Sprzęt medyczny Askle Sante Winncare Automorpho Plus. Askle Sante Winncare Automorpho Plus 20 stron.
ALARMS
REPAIR
AND
The following table shows all the alarms for the device, their triggering conditions, their
priority as well as possible repair.
Description
Alarm LED flashing
and mains present
LED off
+
+ sound signal
Alarm LED flashing
and mains present
LED on
+ sound signal
Fixed alarm LED
+ sound signal
NOTES:
At start up of the compressor, there is an initialization step. In the event of
malfunction, an alarm is indicated by rapid flashing of all the LEDs or by the flashing
of the revision LED or by the flashing of the alarm LED.
When the compressor is disconnected, even voluntarily, there is systematically an
electrical fault alarm to prevent involuntary action.
The triggering conditions in the table above are classified according to their
priority.
The sound and visual alarms are designed to be noticed 2 metres from the
compressor.
WINNCARE FRANCE
Priority
Condition
Electrical fault:
the compressor is
Medium
not powered by
the mains
Compressor
system problem
The compressor
measures a nil
pressure whereas
it pumps
The compressor
Medium
measures an
over-pressure
Inflation
impossible
The compressor
detects a
pneumatic
problem
Initial inflation
Low
impossible
Trigger
Possible repair
Time
frame
1 - Re-connect the compressor
0 min
4- Disconnect the support from
5 - Contact your maintenance
0 to 1
min
1 min
1 min
40
4 - Contact your maintenance
min
1 min
40
min
3 - Contact your maintenance
- 14 -
2 - Check the status of the
electrical network
3 - Replace the fuse
the compressor.
department
1 - Check that the support is
properly connected to the
compressor
2 - Check that the CPR is well
sealed
3 - Remove the patient from
the support
department
1 - Check that the support is
properly connected to the
compressor
2 - Check that the CPR is well
sealed
department