AMCON EQ-6001 Manual do utilizador - Página 11

Procurar online ou descarregar pdf Manual do utilizador para Microscópio AMCON EQ-6001. AMCON EQ-6001 11 páginas. Lensometer

Demarcation of Lens

For demarcation of a spherical lens, rotate the diopter measuring handwheel to focus the lines of
the (3) marker reticule clearly. Lightly move the lens to find the right optical center. Make three-
point printing marks with the printer.
For demarcation of the axis of a sphero-cylindrical lens, rotate the astigmatism axis measuring
handwheel to the required angle. Then rotate the diopter measuring handwheel while turning the
lens to focus the cylindrical marking lines clearly. Lightly move the lens to find the right optical
center. Make three-point printing marks with the printer.
For demarcation prismatic lens base, rotate the astigmatism axis measuring handwheel to the
required angle. Then rotate the diopter measuring handwheel to focus the lines of the (3) marker
reticule clearly. Lightly turn the prism lens so that the middle of the three green marking lines
passes through the center of the (8) eyepiece reticule. Make printing point marks by the use of
the printer.
MAINTENANCE
Do not dismantle the lensometer, as the accuracy will be compromised.
Store the lensometer in a dry, cool, indoor place in order to assure precise working condition.
Prevent the instrument from severe vibration or impact in order to avoid damage or loosening parts.
When changing the bulb, make sure the bulb is cool before touching.
Always keep the instrument clean. Do not touch the surfaces of the optical parts. Dust shall be cleaned
with an absorbent cotton-tipped applicator dipped with a mixture of ethanol and ether.
After each use, clean the instrument and cover it with a dust shield.
Problem
Bulb will not light up.
Cannot see eyepiece reticule when
adjusting the eyepiece readings.
Green marking image cannot be
seen.
Green marking image
cannot be seen clearly.
Deviation of the center
of the green marking image.
Translocation of the printing unit.

COMMON TROUBLE SHOOTING

Reason
Mains are not connected.
Observer has too much
uncorrected refractive error.
Bulb cannot be lighted or
Dust shield is still on.
Dust accumulates
on optical lens.
Lock screws are loose.
Loosening of limit
screw or loosening of
connecting screws.
10
Solution
Connect with the mains.
Wear a refractive correction
when using the lensometer.
Connect with mains or
Take off dust shield.
Clean it using cotton
dipped in cleaning solution.
Remove objective lens shield,
regulate three screws, lock them.
Regulate limit screw & lock
or regulate connecting
screws & lock.