Product Maintenance

1. The maximum capacity of the SD card is 16G. (This is five years' worth of space)
2. During battery charging, the temperature of the device rises. The temperature and humidity sensors are affected. Currently,
measurements are only for reference. After batteries are fully charged and the device cools down, measurements are reliable.
3. Repair – Do not attempt to repair the device or modify the circuitry by yourself. Please contact CO2Meter if the device needs
servicing.
4. Caution – The CO sensor must be replaced every 3 years.
5. Cleaning – Disconnect the power before cleaning. Use a damp cloth, do not use a liquid cleaning agent, such as
benzene, thinner or aerosols.
6. Maintenance – Recommended that the user conducts a comprehensive test and calibration every year to ensure normal
operation of the device.