CO2 Meter CM507 Instruções de utilização - Página 2

Procurar online ou descarregar pdf Instruções de utilização para Detectores de gás CO2 Meter CM507. CO2 Meter CM507 2 páginas. Handheld multi gas detector

Buzzer AL L/AL H:
When AL L is on, the buzzer sounds "Ta Ta Ta"; when AL H is on, the buzzer sounds "Bi Bi Bi".
RS485 communications:
Please use the 4 Pole 3.5mm headphone plug. The headphone plug contacts are defined as below:
For long-distance communication, it is necessary to use a dedicated isolated line to ensure communication
quality. The wiring method is as below:
Power off:
Switch the Power Key (○
) off.
A
Note: When the device is charged with USB, it cannot be turned off.
Calibration:
(1) Single-point calibration
At the same time, long press "▲(○
)" and "▼(○
)" key, CAL is displayed on the screen. Press the "M1" key
G
J
(○
) to select the sensor that needs to be calibrated. Press the "▲(○
F
then press the "Enter" key (○
) to proceed the next setting. Zone1 displays the unadjusted raw value; Zone2
I
displays the adjusted value, and it is also the value of returning to the main screen; Zone1-Zone2=adjustment
amount. After finishing the setting, press and hold the "M1" key (○
F
(2) Two-point calibration
At the same time, long press "▲(○
)" and "▼(○
)" key, CAL is displayed on the screen. And then long press
G
J
the "M2" key (○
), CAL L is displayed on the screen. Next, press the "M1" key (○
H
needs to be calibrated (CO
COO
NH3PM2.5RHAMB) and the corresponding u n i t will blink. Press
2
2
"▲(○
)/▼(○
)" key to adjust the value of the standard gas concentration. Press "Enter" key (○
G
J
CAL H (High gas concentration calibration)/CAL L (low gas concentration calibration). The ppm unit is not
flashing for a moment, indicating that the calibration value has been stored. Long press the "M2" key (○
calibration calculation. At the same time, the screen displays the calibration reading to confirm the accuracy. Short
press the "M1" key (○
) to set the next gas calibration, or long press the "M1" key (○
F
screen.
Clear Two-point calibration:
Set the same calibration value for CAL L and CAL H, and then proceed the calculation. After the calibration is
completed, the original calibration value is cleared.
Rechargeable Batteries
Battery message:
'Battery OK':
'Battery Low':
Measurements are possible
The battery needs to be recharged,
measurements are still possible
Battery installation:
This device is supplied with Li-ion 18650 3.7V rechargeable battery *3 pcs.
Please confirm whether batteries + polarity are Li ion18650 positive bump specification and install batteries in
the correct polarity. If the battery polarity is reversed, the capacity will be insufficient and the battery life will be
shortened.
Battery charging:
The charge lamp (○
(5V/1A USB adapter charger)
Note: During battery charging, the temperature of the device will rise by 5°C~10°C. At this time, the
measurements of temperature and humidity will be affected by temperature rise. Cause an impact on
the accuracy of temperature when charging, please use a fan to blow toward the temperature sensor
(○
) directly in order to get a compensated balance of temperature and humidity between temperature
C
sensor and surrounding area.
Safety Instructions
Warning: Your safety is very important to us. To ensure to use the device correctly and safely, we would like
to draw your attention to read the warning and entire User Manual before using the device. These
are important safety information and should be observed at all times.
1. Please handle the device lightly, do not subject the device to impact or shock.
2. Do not immerse the device in water. Water can cause electric shock, fire or malfunction which may result in
damage.
3. Do not keep the device under the hot and moisture environment. Keep the device away from the heat source
or near water.
4. Please use a standard USB power supply (such as PC's USB port, universal AC adapter with USB port).
Improper power supply can cause serious damage to the device, or result in injury or death to the user.
)/▼(○
)" key to edit the parameters, and
G
J
Product Care
To ensure you receive the maximum benefit from using this device, please observe the follow guidelines.
1. The maximum capacity of the SD card is 16G.
) to return to the main screen.
2. During battery charging, the temperature of the device rises. The temperature and humidity sensors are
affected. At this time, measurements are only for reference. After batteries are fully charged and the device
cools down, measurements are reliable.
3. Repair – Do not attempt to repair the device or modify the circuitry by yourself. Please contact with the local
dealer or a qualified repairman if the device needs servicing.
4. Caution – The CO and NH3 sensors must be replaced every 3 years.
5. Cleaning – Disconnect the power before clean. Use a damp cloth, do not use the liquid cleaning agent, such
as benzene, thinner or aerosols.
) to select the sensor that
F
6. Maintenance – Recommend that the user conducts a comprehensive test and calibration every year to
) to switch
I
Specifications
) for a
H
) to return to the main
F
CO
2
CO
O
2
NH3
PM2.5
PM10
RH
Temperature
Barometer
Operating
Conditions
Storage
Temp.
Power
Supply
Storage
Capacity
Comm.
Interface
'Battery Exhausted':
Alarm
Measurements are not
possible
) lights up while charging; the charge lamp (○
) goes off when charging is completed.
E
E
ensure the normal operation of the device.
Measurement Range
Accuracy
0~9,999ppm
±50ppm or 5% of reading,
(5,001~9,999ppm
whichever is greater (0~5,000ppm)
over range)
0~1,000 ppm
±5%~±10%
0~25%
<2% FS/0.1 mbar
0~100ppm
Performance indicator only
±15% or ±15 μg/m³, whichever is
999 ug/m
3
greater
±3%@25°C (20~80%
RH), others
0~100% RH
±5%
0~50°C
±1°C
50~110
kPa
±0.4kPa
0~40°C (32~104°F), 0~95% RH, non-condensing, with lithium-ion batteries
0~50°C (32~122°F), 0~95% RH, non-condensing, without lithium-ion batteries
-20~60°C (-4~140°F), 0~85% RH, non-condensing, without lithium-ion. batteries
Rechargeable Battery : Li-ion 18650 3.7V *3 (batteries not included)
AC Adapter : 5V±5% >1A, 100~240 VAC, 50/60 Hz
depend on SD card capacity (max.16G SD card)
RS485 ModBus BR19200、N、8、1
80db±5%@10cm
Volume
Weight
305g (without batteries)
Dimension
204.5x91.7x49.6 mm (8.1x3.6x2 inch)
Note: After power-on, it would take 20 minutes for the device to stably measure the temperature and
relative humidity.
EMC/RFI: Readings may be affected if the unit is operated within radio frequency electromagnetic field
strength of approximately 3 volts per meter, but the performance of the instrument will not be
permanently affected.
Sensor warm-up time and logging Interval:
No.
Sensor
Warm-up time
CO
<1 minute
1
2
2
CO
5 minutes
O
<1 minute
3
2
4
NH3
5 minutes
PM2.5
<1 minute
5
PM10
Installation Instruction
1 .
You can put CM-500 on the table (please pull out the Housing Stand
2 . P lease note the following when hanging it on the wall. Choose a suitable location to install
CM-500, fix a screw to the wall.
Package contents
Handheld multi gas detector
Manual
USB 1.5M cable
SD Card
Display Resolution
1ppm
1ppm
0.01%
1 ug/m
3
0.01%
0.01°C
0.1 mmHg
Data logging interval time setting
After powering on and standby time,
Data logging interval
data logging can be operated.
time setting
After power on 5 minutes
2 seconds~60 minutes
After power on 5 minutes
2 seconds~60 minutes
After power on 5 minutes
2 seconds~60 minutes
After power on 5 minutes
2 seconds~60 minutes
After power on 5 minutes
2 seconds~60 minutes
) or hang it on the wall.
P
CO2 Meter, Inc.
131 Business Center Drive,
Ormond Beach, FL 32174
[email protected] www.co2meter.com
Ref.No.:122018