Quick Start Guide
Perform a measurement
1. Acclimatize the device to the surrounding temperature (1-2 hours).
2. Press ON/OFF to switch on the TecPen.
3. The PWR LED will light up green.
4. The TecPen will indicate the battery status and oxygen level with ****%.
5. The TecPen will take 2 minutes to warm up.
6. Connect the gas inlet of the TecPen to the provided filter and connect the sampling needle
onto the filter using the Leur lock.
7. Optional: Connect the tubing to the TecPen. Put the filter onto the free end of the tubing.
Then connect the needle to the filter.
8. Take a septum and place it on the package. Make sure that that the septum is placed on a
place where enough gas is available to make the test.
9. Insert the needle into the gas-filled space to be measured.
10. Press START immediately after piercing the sample.
11. The "MEAS" will appear on the display. The oxygen value, will change continuously during
measurement.
12. If "****" appears your measured value is outside the measuring range.
13. Once the measurement stops changing, press START again to stop the measurement.
14. Press SAVE to save the measured value. The saving process is indicated by "LOG" on
the display. The value is stored in the internal memory.
15. Press ON/OFF again to turn the TecPen off.
For additional settings and features refer to the sections in this manual.
3