preparIng tO run an OptIOnal lIquID qualItY COntrOl test
1. gather the Following Items
• CoaguChek XS Plus Meter
• Container of test strips
• Test Strip Code Chip
• CoaguChek XS Pro PT Control
Solutions
• Diluent Dropper (2)
• Quality Control Code Chip
runnIng an OptIOnal lIquID qualItY COntrOl test
1. remove a test strip
Take a test strip out of the container.
Close the container tightly.
You have 10 minutes to use a test strip
once you remove it from the container.
6. Collect Control
Using the dropper, gently squeeze
Apply a drop of control to the target area.
and hold the bulb. Insert the tip of the
Hold the control drop to the test strip until
dropper into the control solution. Slowly
you hear a beep. The flashing dropper
release the bulb to draw the control up
symbol will disappear. An hourglass will
to the bulb. Avoid creating bubbles in
appear, indicating the test is in progress.
the control.
Return any unused control to the bottle
for additional tests. Reconstituted controls
are good for 30 minutes.
The result appears in less than one
minute.
2. Match Code
The code number on the control
solution and the Quality Control Code
Chip must match.
Each box of controls comes with a
matching code chip. Control code
chips are identified with a "C" in
the number.
2. Insert test strip
Slide the test strip into the test strip
guide in the direction of the arrows
until it stops.
The meter powers on, and the main
menu appears.
7. apply the Control
Record your result as instructed by
your facility.
Note: If during testing the meter displays
an error message, refer to the Error
Messages section of the Operator's
Manual for an explanation and steps on
how to proceed.
Run a level 2 control test with a new test
strip following the same steps.
3. reconstitute Control solution
Open a bottle of lyophilized control.
Cut off the flared tip of the dropper at
the end of the stem.
Hold the diluent dropper vertically by
the stem, not the bulb, to avoid loss of
diluent.
Do not hold the pipette close to
your face.
3. select Control test
Press Control test on the main menu
screen.
An hourglass symbol appears.
The QC Test menu appears.
8. record result
Remove the Quality Control Code Chip
and store it with the remaining controls.
Replace the Test Strip Code Chip.
Roche Diagnostics recommends that you
keep a code chip in the slot to keep the
electrical contacts free of dirt.
Transfer the entire contents of the
Close the bottle.
dropper to the bottle. Do not touch the
Gently swirl the bottle using a circular
dropper to the dried material or to the
motion to dissolve the dried material.
sides of the bottle.
Do not shake or turn the bottle on its
Do not discard the dropper. You will
side or upside down. Avoid creating
need it to apply a drop of control to
bubbles.
the test strip later in the test.
Let the bottle sit undisturbed for at
least a minute to ensure that the dry
material dissolves completely.
4. Insert qC Code Chip
5. select the level
Press new on the QC Test menu.
Select level 1 for running a level one
control.
A code chip symbol with a flashing
arrow appears on the screen.
An hourglass symbol appears.
Remove the Test Strip Code Chip and
A test strip symbol appears with a
insert the Quality Control Code Chip
flashing dropper symbol. You have
with the number facing up until it
180 seconds to apply the quality
snaps into place.
control solution.
The meter stores the code chip
information in memory for future use.
9. remove Control Code Chip
10. Clean up
Place the used strip and control bottles
in an approved biohazard container.
Power the meter off.
Refer to cleaning and disinfecting
procedures in the Operator' s Manual.
ACCU-CHEK, COAGUCHEK and SAFE-T-PRO
are trademarks of Roche.
©2007-2017, Roche Diagnostics.
All rights reserved.
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim, Germany
Made in Germany
Manufactured for and distributed in the USA by:
Roche Diagnostics
9115 Hague Road
Indianapolis, IN 46250-0457
Rx Only
www.coaguchek-usa.com
www.usdiagnostics.roche.com